    Flow Cytometry for Minimal Residual Disease (MRD): Key Factors for Sensitive and Reliable Detection

    |

    Minimal residual disease (MRD) has become an important parameter in the monitoring of patients with hematological malignancies. Detecting a very small population of abnormal cells after treatment can provide information that is not apparent from conventional morphological assessment alone.

    Among the available technologies, multiparameter flow cytometry (MFC) remains an important method for MRD detection because it combines rapid analysis, multiparametric immunophenotyping and broad applicability across hematological diseases.

    However, detecting rare residual cells is not simply a matter of acquiring more events or adding more markers to a flow cytometry panel. Sample quality, immunophenotypic strategy, acquisition, analytical sensitivity and data interpretation all influence the reliability of an MRD result.

    So, what makes a flow cytometry MRD assay reliable?

    What is measurable residual disease (MRD)?

    The term minimal residual disease has increasingly been replaced by measurable residual disease, reflecting the fact that modern techniques can detect disease at very low levels rather than proving that no malignant cells remain.

    In practical terms, MRD refers to the detection of residual malignant cells that remain after treatment, even when the patient may otherwise meet conventional criteria for remission.

    An MRD-positive result indicates that measurable disease has been detected above the validated threshold of the assay. Conversely, an MRD-negative result does not necessarily mean that all malignant cells have been eliminated. It means that residual disease was not detected above the assay’s validated detection capability in the sample analyzed.

    This distinction is particularly important when interpreting highly sensitive assays.

    Why is flow cytometry used for MRD detection?

    Multiparameter flow cytometry can identify abnormal cell populations based on the simultaneous expression of multiple surface and intracellular markers.

    This makes it possible to distinguish leukemic or otherwise abnormal cells from normal hematopoietic populations by evaluating their immunophenotypic characteristics.

    In acute myeloid leukemia (AML), for example, MRD assessment can integrate two complementary strategies:

    • Leukemia-associated immunophenotype (LAIP): tracking an aberrant immunophenotype identified at diagnosis.
    • Different-from-normal (DfN): identifying populations whose immunophenotypic maturation pattern differs from normal hematopoiesis.

    Combining these approaches can help detect both previously identified leukemic populations and emerging abnormal populations. Current ELN recommendations support an integrated LAIP/DfN approach for MFC-MRD in AML.

    How sensitive is flow cytometry for MRD?

    One of the main strengths of flow cytometry MRD is its ability to detect very rare populations.

    Depending on the disease, laboratory, panel, sample quality and analytical strategy, MFC-MRD can reach sensitivities in the range of approximately 10⁻³ to 10⁻⁵, although the achievable sensitivity is not identical across applications or laboratories.

    For example, expert recommendations for adult ALL describe MFC as capable of detecting approximately one leukemic cell among 10,000 nucleated cells (10⁻⁴), although not every laboratory achieves this level.

    This highlights an important point:

    Analytical sensitivity is not simply a property of the flow cytometer. It is a property of the entire assay.

    The instrument, antibodies, sample preparation, number of acquired cells, gating strategy and quality controls all contribute to the final performance.

    1. Sample quality comes first

    Before analyzing rare events, it is essential to ensure that the sample itself is suitable for MRD assessment.

    Factors such as:

    • Cell viability
    • Sample age
    • Sample handling
    • Hemodilution
    • Peripheral blood contamination
    • Cell loss during preparation
    • Insufficient cellularity

    can influence the ability to identify residual disease.

    For AML MRD, the European LeukemiaNet recommends obtaining a first-pull bone marrow aspirate when possible and processing the sample within an appropriate timeframe. Laboratories should also consider strategies for evaluating hemodilution because contamination with peripheral blood can affect the representativeness of the sample.

    Therefore, a technically excellent cytometry analysis cannot compensate for a poor-quality or non-representative sample.

    2. The diagnostic sample can be extremely valuable

    When available, the patient’s diagnostic sample provides an important reference for MRD monitoring. At diagnosis, the leukemic population can be characterized and potential leukemia-associated immunophenotypes (LAIPs) identified. These features can subsequently be tracked after treatment.

    However, leukemic populations can change during therapy. This is one reason why a strategy based exclusively on the original phenotype may not always be sufficient. The integration of LAIP and DfN approaches helps address this challenge by combining patient-specific information with knowledge of normal hematopoietic maturation.

    3. More markers do not automatically mean better MRD detection

    High-parameter flow cytometry allows laboratories to analyze increasingly complex immunophenotypes. But adding markers does not automatically increase the biological resolution of an assay. The value of each marker depends on its ability to contribute useful discriminatory information.

    A robust MRD panel should consider:

    • Antigen expression patterns
    • Fluorochrome selection
    • Antigen density
    • Background staining
    • Population separation
    • Expected normal expression
    • Aberrant expression associated with the disease
    • Compatibility between markers

    The objective is not simply to maximize the number of parameters. The objective is to maximize the ability to distinguish abnormal cells from normal and regenerating populations.

    4. Normal hematopoiesis can look surprisingly abnormal

    One of the most challenging aspects of MRD immunophenotyping is distinguishing residual malignant cells from normal or regenerating hematopoietic populations. After treatment, bone marrow regeneration can produce immunophenotypic patterns that differ from those observed in a healthy sample. Some apparent abnormalities may therefore reflect regenerating or stressed hematopoiesis rather than residual leukemia.

    This is why control samples and knowledge of normal maturation patterns are essential components of reliable MFC-MRD interpretation. The ELN specifically recommends attention to aberrant phenotypes in regenerating bone marrow samples.

    5. Acquisition strategy matters

    When searching for extremely rare populations, the number and quality of acquired events become critical. If the target population represents only a very small fraction of the sample, acquiring an insufficient number of cells can make detection statistically and analytically challenging.

    For AML, ELN recommendations include acquisition of more than 500,000 CD45-positive cells and at least 100 viable cells within the relevant blast compartment for the standard assessment of MFC-MRD negativity. Importantly, acquisition should not be considered separately from quality control.

    Recommended gating and acquisition strategies include parameters such as:

    • FSC versus time
    • Doublet exclusion
    • Viability
    • Appropriate population identification
    • Consistent acquisition settings

    These steps help ensure that the events used for MRD calculations are relevant and technically reliable.

    6. LOD and LOQ are essential for interpreting an MRD result

    One of the most important concepts in flow cytometry MRD is the distinction between detecting a population and reliably quantifying it.

    Limit of detection (LOD)

    The limit of detection (LOD) represents the lowest level at which the assay can reliably distinguish a signal from background.

    Limit of quantification (LOQ)

    The limit of quantification (LOQ) represents the lowest level at which the analyte or target population can be quantified with an acceptable level of performance.

    These parameters are particularly important when dealing with rare populations. A result should therefore not simply state that MRD is “negative” without considering the analytical capability of the assay.

    The ELN MRD recommendations specifically state that LLOD and LLOQ should be calculated to assess MFC-MRD assay performance.

    7. MRD-negative does not necessarily mean zero disease

    This is one of the most important concepts when communicating MRD results. Consider an assay capable of reliably detecting one abnormal cell among 10,000 analyzed cells.

    If no abnormal population is detected, the result does not demonstrate that there are zero leukemic cells in the patient. Instead, it indicates that disease was not detected above the assay’s validated analytical threshold in the sample analyzed. This is why the term “measurable residual disease” is increasingly preferred.

    The distinction is particularly relevant when comparing MRD results obtained using different methodologies or laboratories.

    8. Standardization is critical for reproducible MRD results

    A reliable MRD assay requires more than a good antibody panel. Standardization across the workflow is essential.

    This includes:

    Pre-analytical standardization

    • Sample collection
    • Anticoagulant
    • Sample transport
    • Storage conditions
    • Processing time

    Analytical standardization

    • Sample preparation
    • Antibody combinations
    • Instrument settings
    • Acquisition strategy
    • Quality controls

    Post-analytical standardization

    • Gating strategy
    • Data interpretation
    • Sensitivity calculations
    • Reporting criteria

    The ELN emphasizes that a minimum required set of tubes and fluorochrome combinations is important for harmonized LAIP/DfN MRD detection, analysis and reporting.

    9. Flow cytometry MRD is particularly valuable because of its broad applicability

    MRD can be evaluated using different technologies, including:

    • Multiparameter flow cytometry
    • Quantitative PCR
    • Digital PCR
    • Next-generation sequencing

    Each approach has specific advantages, limitations and areas of applicability.

    For example, molecular approaches can achieve very high analytical sensitivity for suitable molecular targets, but their applicability depends on the biological characteristics of the disease and the availability of an appropriate molecular marker.

    Flow cytometry, in contrast, provides direct cellular and immunophenotypic information and can be applied across a broad range of hematological malignancies.

    For AML, current ELN recommendations use molecular MRD approaches for specific molecularly defined subgroups, while recommending MFC-MRD for patients outside those groups.

    This means that the choice of MRD technology should be based on the disease, biological target, clinical context and validated performance of the assay.

    What makes a reliable flow cytometry MRD assay?

    A robust flow cytometry MRD workflow can be summarized around five essential principles:

    1. High-quality samples

    The sample must be representative, viable and appropriately processed.

    2. Appropriate immunophenotyping

    The panel must identify relevant abnormal populations while distinguishing them from normal and regenerating cells.

    3. Sufficient event acquisition

    Rare-event detection requires acquisition of enough relevant cells to support the desired analytical sensitivity.

    4. Validated analytical performance

    LOD, LOQ and other relevant performance characteristics should be established for the assay.

    5. Standardized analysis and reporting

    The interpretation of MRD results should follow consistent analytical criteria and clearly communicate the sensitivity and quality of the result.

    The future of MRD flow cytometry

    As flow cytometry continues to evolve, MRD analysis is moving towards increasingly standardized, high-dimensional and automated workflows.

    High-quality data will be particularly important for the development of automated analysis approaches. ELN recommendations already highlight the importance of standardized instrument settings, preanalytics and measurements as a foundation for future automated analyses. At the same time, increasing the number of parameters alone will not solve every challenge.

    The future of MRD flow cytometry will depend on the integration of: high-quality reagents + standardized protocols + robust controls + high-quality acquisition + advanced data analysis.

    The ultimate objective is not simply to detect more cells. It is to detect clinically relevant abnormal populations with confidence, reproducibility and a clearly defined analytical sensitivity.

    Conclusion

    Flow cytometry for minimal residual disease detection is a powerful approach for monitoring residual abnormal cell populations in hematological malignancies. However, MRD sensitivity should never be considered solely as an instrument specification. Reliable MRD detection depends on the entire workflow—from sample collection and preparation to antibody selection, acquisition, gating, quality control, analytical validation and reporting.

    As MRD becomes increasingly important in hematological disease monitoring, standardized and reproducible flow cytometry workflows will remain essential for generating results that laboratories and clinicians can interpret with confidence.

    At Immunostep, we develop flow cytometry reagents and solutions for immunophenotyping and oncohematology applications, supporting laboratories working with increasingly demanding analytical workflows. Our portfolio includes antibodies, multicolor cocktails, EuroFlow-related reagents and solutions for applications including PNH, CAR-T and hematological analysis.